File name: Pandora270s1_Warsaw-UW_20240914_L2Fit_fwvt1c6d20231207p1-8.txt File generation date: 20251023T084256.8Z Data description: Level 2 spectral fitting results file Data file version: fwvt1c6p1-8 Local principal investigator: Iwona Stachlewska Network principal investigator: Alexander Cede Instrument type: Pandora Instrument number: 270 Spectrometer number: 1 Processing software version used: BlickP v1.8.75 Instrument operation file used: Pandora270_OF_v3d20230517.txt Instrument calibration file used: Pandora270s1_CF_v6d20231207.txt Level 1 file used: Pandora270s1_Warsaw-UW_20240914_L1_smca1c6d20231207p1-8.txt Full location name: Warsaw Observatory Station, Faculty of Physics, University of Warsaw Short location name: Warsaw-UW Country of location: Poland Location latitude [deg]: 52.2113 Location longitude [deg]: 20.9828 Location altitude [m]: 137 Local noon date: 20240914 Notes on s-number (L1 configuration): Corrections NOT applied although requested by the s-number are latency correction, matrix method stray light correction (replaced by simple method) Data caveats: None First and last pixel inside fitting window: 1624 1775 Nominal wavelengths inside fitting window [nm]: 492.0035 492.1240 492.2444 492.3648 492.4851 492.6055 492.7259 492.8462 492.9665 493.0868 493.2071 493.3273 493.4476 493.5678 493.6880 493.8082 493.9284 494.0485 494.1687 494.2888 494.4089 494.5290 494.6491 494.7691 494.8892 495.0092 495.1292 495.2492 495.3692 495.4891 495.6091 495.7290 495.8489 495.9688 496.0886 496.2085 496.3283 496.4482 496.5680 496.6878 496.8075 496.9273 497.0470 497.1667 497.2864 497.4061 497.5258 497.6454 497.7651 497.8847 498.0043 498.1239 498.2434 498.3630 498.4825 498.6020 498.7215 498.8410 498.9605 499.0799 499.1994 499.3188 499.4382 499.5576 499.6769 499.7963 499.9156 500.0349 500.1542 500.2735 500.3928 500.5120 500.6312 500.7504 500.8696 500.9888 501.1080 501.2271 501.3462 501.4654 501.5844 501.7035 501.8226 501.9416 502.0606 502.1797 502.2986 502.4176 502.5366 502.6555 502.7744 502.8933 503.0122 503.1311 503.2499 503.3688 503.4876 503.6064 503.7252 503.8440 503.9627 504.0814 504.2002 504.3189 504.4375 504.5562 504.6749 504.7935 504.9121 505.0307 505.1493 505.2678 505.3864 505.5049 505.6234 505.7419 505.8604 505.9789 506.0973 506.2157 506.3341 506.4525 506.5709 506.6893 506.8076 506.9259 507.0442 507.1625 507.2808 507.3991 507.5173 507.6355 507.7537 507.8719 507.9901 508.1082 508.2264 508.3445 508.4626 508.5807 508.6987 508.8168 508.9348 509.0528 509.1708 509.2888 509.4068 509.5247 509.6426 509.7605 509.8784 509.9963 --------------------------------------------------------------------------------------- Column 1: Two letter code of measurement routine Column 2: UT date and time for center-time of measurement, yyyymmddThhmmssZ (ISO 8601) Column 3: Fractional days since 1-Jan-2000 midnight for center-time of measurement Column 4: Routine count (1 for the first routine of the day, 2 for the second, etc.) Column 5: Repetition count (1 for the first set in the routine, 2 for the second, etc.) Column 6: Total duration of measurement set in seconds Column 7: Data processing type index Column 8: Solar zenith angle for center-time of measurement in degree Column 9: Solar azimuth for center-time of measurement in degree, 0=north, increases clockwise Column 10: Lunar zenith angle for center-time of measurement in degree Column 11: Lunar azimuth for center-time of measurement in degree, 0=north, increases clockwise Column 12: Pointing zenith angle in degree, absolute or relative (see next column), 999=tracker not used Column 13: Zenith pointing mode: zenith angle is... 0=absolute, 1=relative to sun, 2=relative to moon Column 14: Pointing azimuth in degree, increases clockwise, absolute (0=north) or relative (see next column), 999=tracker not used Column 15: Azimuth pointing mode: like zenith angle mode but also fixed scattering angles relative to sun (3) or moon (4) Column 16: Fitting result index: 0=no error or warning, 1, 2=warning, >2=error Column 17: Number of function evaluations used, 0=linear fitting or fitting not successful or no fitting done Column 18: rms of unweighted fitting residuals, -9=fitting not successful Column 19: Normalized rms of fitting residuals weighted with independent uncertainty, -9=fitting not successful or no uncertainty used Column 20: Expected rms based on independent uncertainty, -9=fitting not successful or no uncertainty given Column 21: Expected normalized weighted rms based on independent uncertainty, -9=fitting not successful or no uncertainty given Column 22: Water vapor slant column amount [moles per square meter], -9e99=fitting not successful Column 23: Independent uncertainty of water vapor slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 24: Structured uncertainty of water vapor slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 25: Common uncertainty of water vapor slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 26: rms-based uncertainty of water vapor slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 27: Water vapor effective temperature [K] Column 28: Independent uncertainty of water vapor effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 29: Structured uncertainty of water vapor effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 30: Common uncertainty of water vapor effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 31: Effective water vapor fitting wavelength [nm], -9=fitting not successful Column 32: Diffuse correction applied before fitting at effective fitting wavelength for water vapor [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 33: Nitrogen dioxide slant column amount [moles per square meter], -9e99=fitting not successful Column 34: Independent uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 35: Structured uncertainty of nitrogen dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 36: Common uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 37: rms-based uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 38: Nitrogen dioxide effective temperature [K] Column 39: Independent uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 40: Structured uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 41: Common uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 42: Effective nitrogen dioxide fitting wavelength [nm], -9=fitting not successful Column 43: Diffuse correction applied before fitting at effective fitting wavelength for nitrogen dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 44: Ozone slant column amount [moles per square meter], -9e99=fitting not successful Column 45: Independent uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 46: Structured uncertainty of ozone slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 47: Common uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 48: rms-based uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 49: Ozone effective temperature [K] Column 50: Independent uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 51: Structured uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 52: Common uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 53: Effective ozone fitting wavelength [nm], -9=fitting not successful Column 54: Diffuse correction applied before fitting at effective fitting wavelength for ozone [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 55: Molecular scattering air mass factor used for molecular scattering subtraction before the fitting Column 56: Estimated uncertainty of molecular scattering air mass factor, -9=molecular scattering was not subtracted before the fitting Column 57: Lower limit used for wavelength scaling [nm] Column 58: Upper limit used for wavelength scaling [nm] Column 59: Order of smoothing polynomial Column 60: Smoothing polynomial coefficient, order 0 Column 61: Independent uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 62: Structured uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 63: Common uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 64: rms-based uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 65: Smoothing polynomial coefficient, order 1 Column 66: Independent uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 67: Structured uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 68: Common uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 69: rms-based uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 70: Mean value of measured data inside fitting window [same units as measurements] Column 71: Wavelength change polynomial coefficient, order 0 Column 72: Independent uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 73: Structured uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 74: Common uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 75: rms-based uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 76: L1 based wavelength change coefficient, order 0, -9=no L1 wavelength change determination Column 77: L1 based wavelength change coefficient, order 1, -9=no L1 wavelength change determination Column 78: L2Fit data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality Column 79: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ1 limit, 0=L1 data quality above 0, 1=Spectral fitting was not successful, 2=Wavelength shift too large, 3=Normalized rms of fitting residuals weighted with independent uncertainty too large Column 80: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ2 limit (same parameters as for DQ1) Column 81: L1 data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality Column 82: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ1 limit, 0=Saturated data, 1=Too few dark counts measurements, 2=No temperature given or effective temperature too different from the reference temperature, 3=Dark count too high, 4=Unsuccessful dark background fitting, 5=The dark count differs significantly from the dark map for too many pixels, 6=Absolute value of estimated average residual stray light level too high, 7=Although attempted, no wavelength change could be retrieved, 8=Absolute value of retrieved wavelength shift too large, 9=Retrieved wavelength shift differs too much from the shift predicted by the effective temperature Column 83: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ2 limit (same parameters as for DQ1) Column 84: Atmospheric variability [%], 999=no atmospheric variability was determined Column 85: Wavelength effective temperature [°C], 999=no effective temperature given Column 86: Estimated average residual stray light level [%] (only valid for stray light correction methods 2 and higher) Column 87: Retrieved wavelength shift from L1 data [nm], -9=no wavelength change determination Column 88: Retrieved total wavelength shift [nm], -9=no wavelength change fitting Column 89: Number of bright count cycles Column 90: Number of dark count cycles Column 91: Effective position of filterwheel #1, 0=filterwheel not used, 1-9 are valid positions Column 92: Effective position of filterwheel #2, 0=filterwheel not used, 1-9 are valid positions Column 93: Sum over 2^i, 0=spectra were interpolated in time, 1=spectra are corrected for off-target signal Column 94: Integration time [ms] Column 95: Spectrometer control temperature [°C], 999=no temperature signal Column 96: Auxiliary spectrometer temperature [°C], 999=no temperature signal Column 97: Temperature in head sensor [°C], 999=no temperature signal Column 98: Temperature at electronics board 1 [°C], 999=no temperature signal Column 99: Temperature at electronics board 2 [°C], 999=no temperature signal Column 100: Humidity in head sensor [%], -9=no humidity signal Column 101: Pressure in head sensor [hPa], -9=no pressure signal Columns 102-253: Unweighted slant column residuals for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting Columns 254-405: Normalized slant column residuals weighted with independent instrumental uncertainty for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting, 0=no independent instrumental uncertainty was given --------------------------------------------------------------------------------------- SQ 20240914T102647.2Z 9023.435269 242 1 54.49 2 49.09 178.44 135.94 70.13 0.00 1 0.00 1 0 15 3.367e-03 3.336e-03 7.636e-04 7.620e-04 4.121402e+03 4.582185e+01 3.113330e+02 -6.000000e+00 1.796003e+02 279.60 -2.00 20.00 -2.00 505.27 0.00e+00 4.360760e-04 2.044619e-05 3.048567e-05 -6.000000e+00 9.230141e-05 279.60 -2.00 20.00 -2.00 500.43 0.00e+00 1.046811e-01 1.478130e-02 2.646946e-02 -6.000000e+00 6.539167e-02 228.70 -2.00 3.84 -2.00 502.03 0.00e+00 1.524 0.000 489.00 513.00 1 8.0640e+00 1.1309e-03 3.2189e-03 -6.0000e+00 5.0078e-03 5.4360e-03 2.9096e-04 6.9232e-04 -6.0000e+00 1.2965e-03 4.8819e-04 8.9737e-04 6.7788e-05 4.1259e-05 -6.0000e+00 3.0889e-04 5.5901e-04 -2.5127e-05 10 0 0 10 0 0 100.00 25.86 1.81 0.00056 0.00678 89 9 1 2 0 612.210 20.00 25.86 24.76 34.75 33.75 2.8 959.91 -275 -612 -144 -274 8 232 583 596 724 586 302 252 -159 -42 -312 -359 -42 153 161 -378 118 237 265 167 3 7 381 293 -22 -217 -481 -245 -375 154 235 -397 15 -51 -14 -130 -208 180 12 -62 348 274 267 479 389 162 -390 -484 -759 -758 -406 -298 -314 71 -24 -37 -59 236 256 -173 210 172 79 518 212 -439 -418 97 -30 -176 -23 6 -308 -209 -340 -250 -303 -108 -338 -453 -338 -30 -385 -233 81 -150 43 -103 321 551 265 438 299 649 -412 -536 -870 -756 -639 -183 198 67 -135 -148 -299 -219 -34 -9 121 208 179 -2 512 176 262 401 102 -66 65 307 388 148 -9 492 78 420 586 56 -121 41 -629 -242 -84 146 -30 171 26 96 409 179 314 855 41 -114 -561 -402 -166 -750 -266 -598 -139 -276 9 236 605 628 763 612 312 257 -161 -43 -316 -364 -43 154 167 -401 126 252 281 177 3 7 407 310 -23 -218 -488 -247 -382 155 246 -415 15 -54 -15 -134 -214 184 13 -65 356 283 275 504 403 164 -387 -481 -741 -737 -399 -293 -318 73 -24 -38 -60 242 265 -174 217 176 80 511 208 -426 -414 95 -30 -171 -23 6 -302 -201 -332 -241 -292 -104 -332 -433 -326 -29 -380 -230 79 -147 42 -102 323 559 269 437 292 618 -398 -505 -809 -703 -607 -174 195 66 -132 -148 -295 -214 -33 -9 118 206 178 -2 514 179 266 403 101 -64 64 301 382 142 -9 488 75 415 577 55 -117 38 -595 -227 -80 142 -29 166 26 96 407 180 315 855 41 -112 -540 -390 -160 -711 SQ 20240914T103200.0Z 9023.438889 246 1 55.36 2 49.08 180.17 135.21 71.36 0.00 1 0.00 1 0 22 3.595e-03 3.561e-03 8.834e-04 8.815e-04 5.700165e+03 5.459870e+01 4.157042e+02 -6.000000e+00 1.917372e+02 279.60 -2.00 20.00 -2.00 505.27 0.00e+00 4.426318e-04 2.373434e-05 3.537840e-05 -6.000000e+00 9.855334e-05 279.60 -2.00 20.00 -2.00 500.47 0.00e+00 1.779575e-01 1.718482e-02 3.773462e-02 -6.000000e+00 6.996136e-02 228.70 -2.00 3.84 -2.00 502.05 0.00e+00 1.524 0.000 489.00 513.00 1 8.3495e+00 1.3152e-03 3.9730e-03 -6.0000e+00 5.3538e-03 7.0911e-03 3.3870e-04 9.6458e-04 -6.0000e+00 1.3933e-03 3.6404e-04 1.1196e-03 7.7507e-05 5.1590e-05 -6.0000e+00 3.3048e-04 1.3160e-04 -2.6746e-05 11 1 0 11 32 0 99.92 25.86 1.69 0.00013 0.00790 27 3 1 2 0 2050.020 20.00 25.86 25.03 34.75 33.75 2.9 961.60 900000 -639 -324 -376 163 10 457 479 876 601 566 93 -110 70 -466 -501 -198 231 291 -423 82 60 138 156 169 117 349 289 126 -132 -525 -355 -236 113 -82 55 38 -295 53 -471 -371 288 256 -35 498 276 382 633 334 182 -349 -530 -674 -693 -369 -265 -66 -85 -50 -183 56 315 153 15 352 280 -328 637 241 -729 -532 -37 -74 -301 -222 99 -196 -179 -409 -445 -266 -93 -330 -435 -182 -117 -484 92 -63 -138 74 -47 430 526 143 484 246 607 -463 -587 -753 -809 -567 -55 96 90 -77 -81 -180 -121 -248 -166 -8 140 240 -57 471 249 404 469 357 -180 -152 409 612 199 47 492 -4 306 332 159 -98 86 -718 -243 65 446 109 127 93 163 585 313 162 555 4 -410 -315 -604 -773 -171 900000 -620 -323 -377 167 11 472 502 915 623 597 97 -114 71 -469 -506 -202 238 298 -446 87 63 145 164 179 124 369 304 130 -136 -529 -356 -238 117 -86 59 40 -310 54 -483 -379 292 262 -36 522 291 402 659 343 184 -354 -522 -652 -667 -360 -267 -66 -87 -52 -188 57 320 157 16 360 284 -328 642 241 -725 -521 -36 -73 -299 -223 100 -196 -175 -395 -425 -252 -89 -319 -424 -179 -116 -472 92 -63 -138 72 -46 424 525 142 471 242 582 -435 -557 -704 -759 -524 -52 93 90 -76 -80 -175 -116 -240 -163 -8 141 242 -57 478 247 402 478 359 -178 -148 393 590 194 46 478 -4 296 322 153 -93 82 -689 -231 62 424 106 125 93 164 571 308 161 547 4 -410 -307 -578 -737 -164 SQ 20240914T121701.2Z 9023.511820 308 1 52.01 2 53.78 213.37 119.92 92.80 0.00 1 0.00 1 0 15 6.061e-03 6.058e-03 1.585e-03 1.580e-03 4.441191e+03 9.613660e+01 3.424488e+02 -6.000000e+00 3.232682e+02 279.60 -2.00 20.00 -2.00 505.27 0.00e+00 2.852275e-04 4.243362e-05 2.816368e-05 -6.000000e+00 1.661241e-04 279.60 -2.00 20.00 -2.00 500.40 0.00e+00 3.295625e-01 3.075579e-02 2.679752e-02 -6.000000e+00 1.176619e-01 228.69 -2.00 3.84 -2.00 502.00 0.00e+00 1.688 0.000 489.00 513.00 1 9.2509e+00 2.3530e-03 3.3990e-03 -6.0000e+00 9.0111e-03 8.2059e-03 6.0498e-04 7.2753e-04 -6.0000e+00 2.3322e-03 1.4341e-04 8.8265e-04 1.4005e-04 4.3551e-05 -6.0000e+00 5.5595e-04 -1.2940e-03 7.7627e-06 12 8 8 10 0 0 99.14 25.88 1.65 -0.00129 0.00483 13 1 2 2 0 4000.000 20.00 25.88 26.63 34.75 33.75 2.8 967.15 -93 -445 -125 -583 380 -215 753 590 1829 827 509 157 -439 560 -787 -538 49 470 320 -923 170 90 115 -153 -558 79 487 100 300 449 -1064 -142 74 264 -87 -960 -30 -627 -79 -288 -211 1041 299 -566 832 526 57 1202 384 63 -490 -774 -493 -504 -435 -578 -59 -254 -290 -781 -220 437 -377 -624 620 170 -178 1202 589 -1218 -649 427 548 -136 -337 74 -207 -231 -244 242 208 180 -233 -493 -739 -456 -1158 215 -14 -298 246 -592 525 599 -38 462 323 1340 -560 -891 -792 -508 -1030 406 377 333 -284 -259 -117 445 -327 86 718 110 542 -898 207 -215 224 654 -188 -223 -249 425 436 197 -447 613 -713 14 370 363 -25 617 -1113 -46 291 1093 -293 87 -66 189 1188 -133 94 2573 -373 -576 -792 169 185 -1716 -88 -427 -123 -598 398 -221 781 621 1907 862 524 160 -443 575 -812 -555 51 473 331 -986 181 96 123 -163 -600 84 519 106 311 448 -1101 -145 76 266 -91 -1040 -32 -665 -82 -297 -217 1058 309 -596 874 558 61 1251 396 64 -485 -779 -485 -494 -433 -568 -60 -261 -300 -809 -226 447 -390 -649 634 173 -179 1171 570 -1182 -618 420 543 -131 -329 72 -201 -220 -239 234 200 175 -219 -468 -710 -442 -1141 209 -13 -290 239 -585 520 601 -38 455 312 1249 -541 -841 -736 -472 -982 378 366 331 -285 -257 -114 427 -318 85 712 111 549 -926 212 -213 221 668 -190 -222 -244 411 424 193 -441 598 -704 13 360 351 -23 574 -1043 -43 272 1039 -288 87 -67 183 1194 -130 92 2578 -365 -554 -778 162 175 -1675 SQ 20240914T123946.2Z 9023.527618 319 1 50.52 2 55.87 219.84 116.56 96.93 0.00 1 0.00 1 0 15 3.378e-03 3.359e-03 8.249e-04 8.234e-04 4.519567e+03 4.976309e+01 3.325949e+02 -6.000000e+00 1.801920e+02 279.59 -2.00 20.00 -2.00 505.27 0.00e+00 5.081913e-04 2.214760e-05 3.586622e-05 -6.000000e+00 9.260112e-05 279.59 -2.00 20.00 -2.00 500.45 0.00e+00 1.643672e-01 1.596025e-02 3.056400e-02 -6.000000e+00 6.559275e-02 228.68 -2.00 3.84 -2.00 502.04 0.00e+00 1.777 0.000 489.00 513.00 1 8.1125e+00 1.2216e-03 3.7184e-03 -6.0000e+00 5.0233e-03 5.8949e-03 3.1394e-04 7.7777e-04 -6.0000e+00 1.3002e-03 4.4614e-04 5.6247e-04 7.2873e-05 4.3400e-05 -6.0000e+00 3.0990e-04 -8.2548e-04 -1.4499e-06 10 0 0 10 0 0 99.33 25.90 1.74 -0.00083 0.00308 21 2 1 2 0 2405.570 20.00 25.90 27.44 34.75 33.50 2.8 971.74 -328 -682 -335 -251 195 130 677 441 930 475 446 256 -278 131 -345 -380 -191 -21 143 -250 99 191 18 72 32 341 300 318 38 -151 -600 -285 -257 73 182 -426 57 -131 -76 -356 -134 254 240 -98 454 376 365 604 301 176 -421 -441 -591 -639 -479 -396 -107 -145 -27 -264 130 274 153 -155 382 197 71 485 77 -698 -412 -14 -100 -122 14 223 -207 -203 -385 -211 -402 -5 -371 -253 -186 -158 -413 -156 -27 -138 78 22 515 270 195 224 257 554 -202 -482 -791 -701 -622 -97 361 39 -74 -90 -297 -199 -130 -162 228 5 241 -30 382 257 376 457 249 -74 62 294 487 92 -11 459 -245 215 286 88 -39 98 -474 -304 92 327 -12 82 27 183 268 179 194 1086 -159 -254 -450 -217 -180 -768 -322 -655 -329 -256 199 136 715 459 966 491 468 266 -285 131 -345 -380 -193 -21 150 -262 104 201 19 75 34 360 316 332 39 -154 -601 -283 -258 75 188 -454 61 -137 -79 -364 -136 258 245 -101 474 396 383 627 309 177 -426 -432 -568 -631 -466 -397 -107 -148 -28 -270 131 278 156 -160 389 199 71 488 76 -692 -401 -14 -98 -121 14 224 -206 -199 -370 -200 -382 -5 -359 -246 -183 -157 -415 -156 -27 -138 78 22 510 271 195 221 256 536 -191 -459 -744 -661 -579 -93 350 39 -74 -89 -289 -192 -126 -159 227 5 245 -30 380 258 377 455 252 -74 61 285 474 91 -11 450 -242 210 280 85 -37 95 -456 -291 89 314 -12 81 27 180 264 178 194 1074 -158 -248 -441 -208 -171 -743 SQ 20240914T124752.3Z 9023.533244 323 1 31.95 2 56.68 222.06 115.37 98.37 0.00 1 0.00 1 0 15 2.833e-03 2.811e-03 1.388e-04 1.387e-04 3.285495e+03 8.109810e+00 2.463461e+02 -6.000000e+00 1.510882e+02 279.59 -2.00 20.00 -2.00 505.29 0.00e+00 5.184384e-04 3.729852e-06 3.269289e-05 -6.000000e+00 7.766119e-05 279.59 -2.00 20.00 -2.00 500.61 0.00e+00 4.112142e-02 2.675453e-03 2.316625e-02 -6.000000e+00 5.513163e-02 228.68 -2.00 3.84 -2.00 502.18 0.00e+00 1.815 0.000 489.00 513.00 1 4.2047e+00 2.0453e-04 3.3226e-03 -6.0000e+00 4.2189e-03 -1.0233e-02 5.2881e-05 5.9501e-04 -6.0000e+00 1.0980e-03 2.2368e-02 9.6226e-04 1.2398e-05 3.4012e-05 -6.0000e+00 2.6043e-04 1.4812e-03 -3.5529e-05 12 1 1 12 1 1 99.89 25.90 1.67 0.00148 0.00816 400 71 1 2 0 79.650 20.00 25.90 27.89 34.75 33.75 2.8 973.18 900000 -578 -294 -24 10 317 477 499 416 309 305 267 -88 -241 -209 -337 -256 -9 134 6 98 210 79 167 252 207 409 454 -81 -428 -277 -357 -470 52 202 -29 121 92 -87 -209 -113 -112 -1 94 224 299 257 308 247 39 -309 -453 -684 -634 -389 -250 -176 -14 34 40 106 198 264 221 304 139 74 112 -11 -295 -396 -251 -203 -104 89 -27 -195 -99 -206 -401 -390 -236 -320 -323 -64 -54 -133 -171 -115 -175 -116 13 341 335 264 400 407 117 -188 -454 -640 -689 -501 -223 71 -120 -40 -114 -387 -233 -104 -122 6 184 232 314 495 566 467 317 302 92 2 185 239 22 159 224 90 329 404 -46 -65 14 -348 -340 -25 -44 -114 133 193 77 201 277 329 396 247 -68 -340 -397 -426 -414 900000 -554 -286 -23 10 324 491 514 431 317 312 269 -88 -242 -208 -336 -255 -9 138 7 101 217 82 173 261 215 424 467 -83 -431 -274 -356 -468 52 207 -30 125 95 -88 -213 -115 -114 -1 96 230 308 265 318 252 40 -308 -443 -665 -613 -382 -248 -177 -14 35 40 107 201 269 227 310 140 74 111 -11 -289 -387 -245 -200 -102 89 -26 -194 -96 -200 -386 -375 -228 -313 -314 -63 -53 -132 -169 -113 -174 -115 13 343 337 266 400 405 114 -183 -431 -606 -655 -479 -218 71 -120 -40 -113 -384 -228 -103 -121 6 185 235 316 501 576 477 321 304 92 2 182 236 22 158 223 89 326 400 -45 -63 14 -337 -327 -24 -43 -112 132 192 78 203 281 331 399 250 -67 -335 -387 -413 -406 SQ 20240914T125318.7Z 9023.537022 327 1 49.14 2 57.25 223.54 114.58 99.34 0.00 1 0.00 1 0 22 3.783e-03 3.753e-03 9.851e-04 9.829e-04 4.833809e+03 6.019855e+01 3.600201e+02 -6.000000e+00 2.017505e+02 279.59 -2.00 20.00 -2.00 505.27 0.00e+00 4.879574e-04 2.639097e-05 3.544244e-05 -6.000000e+00 1.036786e-04 279.59 -2.00 20.00 -2.00 500.42 0.00e+00 2.135526e-01 1.910901e-02 3.135464e-02 -6.000000e+00 7.343474e-02 228.68 -2.00 3.84 -2.00 502.02 0.00e+00 1.842 0.000 489.00 513.00 1 8.3918e+00 1.4622e-03 3.8382e-03 -6.0000e+00 5.6240e-03 6.5876e-03 3.7546e-04 8.1416e-04 -6.0000e+00 1.4556e-03 3.3302e-04 6.2274e-04 8.6674e-05 4.6690e-05 -6.0000e+00 3.4697e-04 -1.7431e-03 9.2277e-06 10 0 0 10 0 0 99.97 25.90 1.69 -0.00174 0.00258 21 2 1 2 0 2339.980 19.99 25.90 28.08 34.75 33.75 2.8 973.85 -447 -463 -171 -363 271 27 585 518 1004 608 445 234 -319 35 -279 -495 -287 194 286 -400 214 119 303 -77 133 -14 175 504 126 -47 -589 -294 -174 138 77 -550 150 -192 -106 -218 -342 422 78 -278 451 421 90 754 416 171 -521 -620 -658 -622 -533 -366 -161 -141 179 -211 132 369 61 -8 444 299 -113 618 448 -770 -653 18 -53 -115 -235 -79 -326 -260 -245 -32 -219 12 -264 -295 -322 -271 -521 -21 -54 -7 72 -223 578 405 58 416 376 762 -369 -683 -845 -813 -718 206 111 150 -101 -11 -295 -49 -187 14 185 -37 274 -146 356 258 177 506 91 -144 -31 410 390 124 -100 347 -237 249 535 92 134 75 -779 -265 61 589 -1 -51 120 185 605 213 201 912 -37 -268 -494 -363 -46 -851 -432 -450 -170 -364 278 29 604 543 1049 631 456 245 -330 35 -281 -501 -292 201 294 -423 226 126 320 -81 141 -15 186 530 131 -48 -595 -295 -176 139 80 -590 159 -202 -110 -223 -349 429 80 -291 474 445 95 785 428 173 -532 -614 -639 -601 -523 -370 -162 -144 183 -217 134 377 62 -8 455 303 -114 605 449 -770 -643 18 -53 -114 -229 -77 -317 -256 -237 -31 -209 12 -257 -290 -318 -263 -510 -20 -53 -7 70 -219 573 407 58 410 362 738 -351 -635 -775 -747 -673 198 108 149 -101 -11 -288 -47 -182 14 185 -37 279 -145 353 259 178 503 93 -144 -30 398 381 122 -99 341 -233 243 522 89 127 70 -731 -247 57 564 -1 -51 117 181 595 211 200 902 -37 -262 -486 -350 -44 -827 SQ 20240914T131305.5Z 9023.550758 339 1 12.23 2 59.44 228.74 111.70 102.83 0.00 1 0.00 1 0 22 3.292e-03 3.263e-03 1.372e-03 1.370e-03 4.161098e+03 8.196442e+01 3.110844e+02 -6.000000e+00 1.755455e+02 279.59 -2.00 20.00 -2.00 505.28 0.00e+00 5.149468e-04 3.681455e-05 3.317744e-05 -6.000000e+00 9.025328e-05 279.59 -2.00 20.00 -2.00 500.54 0.00e+00 1.332835e-01 2.652546e-02 2.754262e-02 -6.000000e+00 6.415728e-02 228.68 -2.00 3.84 -2.00 502.12 0.00e+00 1.959 0.000 489.00 513.00 1 7.5151e+00 2.0284e-03 3.7198e-03 -6.0000e+00 4.9084e-03 -4.3447e-03 5.2360e-04 7.1976e-04 -6.0000e+00 1.2797e-03 7.9320e-04 3.3507e-03 1.2179e-04 4.1962e-05 -6.0000e+00 3.0287e-04 9.7121e-04 -1.3445e-04 12 1 1 12 1 1 95.36 25.91 1.66 0.00097 0.02421 4 2 1 2 0 2445.440 20.00 25.91 28.16 34.75 33.75 2.7 973.35 900000 -480 -462 -232 102 358 596 477 824 310 416 112 -167 22 -305 -560 -481 196 177 -401 93 93 13 80 37 176 475 291 72 -213 -626 -107 -235 114 259 -207 78 44 124 -349 -249 107 63 -7 497 244 276 351 302 177 -339 -372 -665 -691 -356 -17 -24 -76 -8 -32 108 161 238 108 154 377 255 437 10 -591 -508 -275 -6 -127 -95 153 -156 -271 -496 -461 -412 -341 -383 -446 -295 -57 -135 -328 65 -216 68 188 255 143 145 427 514 403 -300 -640 -560 -814 -509 -291 251 -17 22 -275 -182 -86 44 101 156 135 292 -42 579 253 222 216 303 187 109 285 194 251 144 442 133 410 342 -15 -387 -19 -413 -266 120 295 -365 360 60 -28 356 -173 30 871 113 114 -373 -615 -257 -494 900000 -463 -447 -233 102 368 608 495 853 320 431 116 -169 22 -309 -556 -482 200 180 -420 97 98 14 84 39 186 499 304 74 -217 -622 -106 -234 116 268 -215 81 45 128 -358 -253 109 65 -7 509 252 285 358 311 179 -342 -364 -654 -677 -346 -17 -24 -78 -8 -33 110 165 246 109 159 383 255 430 10 -587 -497 -267 -6 -126 -93 150 -152 -266 -481 -442 -394 -328 -373 -437 -291 -56 -133 -324 63 -212 67 186 256 142 144 427 503 395 -288 -599 -529 -773 -480 -283 248 -17 22 -271 -180 -84 43 101 155 136 295 -42 590 255 224 222 305 185 107 282 189 246 141 433 130 400 334 -15 -379 -19 -395 -253 116 290 -360 358 61 -28 354 -173 30 871 113 115 -367 -593 -252 -479